Cylinders Gas Delivery System : PISS
-pin index safety system (PISS) connection to the anesthetic machine
-cylinders are a high-pressure source of medical gas which is attached to the machines hanger yoke assembly
hanger yoke assembly includes:
-index pins (PISS)
-washer
-gas filter
-check valve
Cylinders are under large values of pressure therefore need to be measured and regulated:
Bourdon pressure gauge: allows measurement of pressure within the cylinder
Pressure regulator: pressure within the cyliner is regulated to less than 50 psig to ensure safety
Hospital Pipeline Gas Delivery System DISS
-diameter index safety system (DISS) connection to the anesthetic machine
-pipeline gases supplied at a pressure between 45 - 55 psig therefore does not need a pressure regulator
-pipeline gases flow through a pressure gauge and check
valve then combine and share a common gas pathway w/ cylinder gases
Oxygen-Pressure-Failure-Device
-nitrous oxide and air supply travels directly to the flow meter
-oxygen passes through a pressure-failure device, oxygen flush valve, and ventilator power outlet
-if the pressure of oxygen < 25 psig then the fail safe valve automatically closes
-fail safe valve automatically closes the gas lines to nitrous oxide and air
-fail safe valve closes gas lines to other gases other than oxygen in order to prevent hypoxic gas mixtures
-fail safe valve does not
ensure safety from other causes of hypoxic mixtures since it is measuring pressure vs concentration of 02
Oxygen Flush Valve
-allows for high flow of oxygen approx. 35 - 75 liters per minute directly into the common gas outlet
-potential for barotrauma to the lung due to high flow rates with no pressure regulator
-therefore extreme caution must be used when using oxygen flush valve as the patient is connected to the breathing circuit
Flow Control Valves
-determines the gas flow rate
-counter clockwise direction increases the gas flow rate
-valves are touch and color coded to allow better identity for each gas flow rates being controlled
Flow Meters
-measure gas flow rate
-constant-pressure variable-orifice flowmeters: are the ones obtained on the anethetic machine
-indicator of measured flow rates may be: ball, bobbin, or float
Flowmeters are callibrated for specific gases which are affected by gas properties with differing patterns of flow
low laminar flows: affected by gas viscosity
high turbulent flows: affected by gas density
Malfunction of flowmeters may be due to:
-dirt in the flow tune
-nonvertical alignment of the tube
-sticking of the float at the top of the tube
-hypoxic mixtures are possible if a leack develops within or downstream of the 02 flowmeter
-risk of hypoxic mixture is reduced if the 02 flow meter is positioned downstream from other gases
Spirometers
-measures
tidal volume delivered to the patients
-many spirometers use a turbine principle to measure both minute ventilation and tidal volume
-pneumotachograph is a form of fixed-orifice flowmeter that may function as a spirometer
Inaccurate measurements of exhaled tidal volumes can be from:
-circuit leaks
-disconnection
-ventilator malfunction
-inertia, friction, and water condensation in wright respirometers
Significant differences between the volume of gas delivered to the circuit and the volume of gas reaching the patient may be due to:
-long compliant breathing tubes
-rapid respiratory rates
-high airway pressures
Breathing Circuit Pressure Gauge
-generally measures the pressure within the expiratory and inspiratory unidirectional valves
-generally reflects airway pressure
Increase in airway pressure may indicate:
-decrease
in pulmonay compliance
-increas in tidal volume
-obstruction within the breathing circuit
Decrease in airway pressure may indicate
-increase in pulmonary compliance
-decrease in tidal volume
-possible leak within the breathing circuit
peak airway pressure: = dynamic compliance which involves movement of gas flow (ex. mucous plug/ kinking of ETT)
plateau airway pressure:= static compliance which does not involve moment of gas flow ( ex. decreased lung compliance)
-generally in normal ventilation in a normal lung: peak airway pressure is equal to plateau pressure
peak airway pressure may be slighty higher than plateau pressure as wel
l
Increase in peak airway pressure and plateau pressurecan be from:
-decreased lung compliance
-increased tidal volume
Increase in peak airway pressure and no change in plateau pressure can be from:
-increased airway resistance
-increased inspiratory gas flow rate
Vaporizers
Four types:
copper kettle vaporizer
measured flow vaporizer
agent specific vaporizer
variable bypass vaporizer
-vaporizes liquid states of anesthetic agents into a gaseous state of anethetic agent
-vapor pressure of anethetic agent: created by the collisional forces of anesthetic molecules upon the walls of the container
-high temperature is associated with high vapor pressure
Copper Kettle Vaporizer
-considered a measured-flow or flowmeter-controlled vaporizer
-copper used because of a high specific heat and high thermal conductivity
-mainly used in veterinarian anesthesia
-dedicated thorpe tube flow meter determines the amount of carrier gas passes through the vaporizer
-vaporizer-circuit flow-control valve divides the vaporizer circuit from the standard oxygen and nitrous oxide flowmeters
-relationship between copper kettle flow, total gas flow, and anesthetic concentration which exits
1ml liquid anesthetic = approximately 200 ml anesthetic vapor after having passed through the vaporizer
Measured Flow Vaporizer
ex. copper kettle vaporizer
vapor output:
(carrier gas to vaporizer) x (vapor pressure of anesthetic agent) /(barometric pressure) -vapor pressure of anesth. agent
= (CG) x (VP) / BP - VP
anesthetic concentration
vapor output / total gas flow within the circuit
Agent Specific Vaporizer
-delivers a constant concentration of anesthetic agent despite temperature changes or flow through the vaporizer
-also known as variable-bypass vaporizer
-counterclockwise motion allows for an increasing percentage of gas flow to enter into the vaporizing chamber vs bypass chamber
-gases from vaporizing chamber and bypass chamber combine together and mix with gases from co2 absorber to form fresh gases
-vaporizers are agent specific therefore must not be mixed accidently with a wrong and noncorresponding anesthetic agent
-excessive tilting of vaporizer may lead to high anesthetic concentrations which may be hazardous to the patient
"pumping effect" may occur at low flows during positive pressure ventilation and entails a reverse flow of gas into the vaporizer
Tec 6
-heat blended vaporizer (ex.desflurane vaporizer)
-electrically heated to 39 degrees C
-vapor pressure of 2 atm
-no FGF through desflurane sump
-pure desflurane vapors mixes with fresh gas just before exiting the vaporizer
-dial control and FGF rate determine the amount of desflurane vapor is released
-does not automatically compensate for in elevation ex. high elevations: anesthesiologist must manually increase the conc. dial
Variable-bypass Vaporizer
-recommended location outside the circle system
-vaporizer between the flowmeter and common gas outlet allows less chances of dramatic inc in conc. with 02 flush valve use
recommended order of vaporizers based on vapor pressure and potency of anesthetic agents
(upstream to downstream): desflurane, methoxyflurane, enflurane, sevoflurane, isoflurane, halothane
Electronically Controlled Vaporizers
-may be integrated with an electronic flow-c
ontrol
Flow-control helps to separate the gas flow into:
bypass flow
liquid chamber flow : agent-specific cassette (Aladin cassette) which vaporizes the anesthetic agent
-the flow leaving the saturated anesthetic liquid chamber combines with the bypass flow to join before leaving the fresh gas outlet -altering the ratio between the ratio between the bypass flow and liquid chamber flow changes the concentration of the anesthetic
-software determines the fresh gas concentration of the anesthetic agent depending on the output pulses from the agent wheel
-liquid chamber flow is calculated dependant on desired fresh gas concentrations and determined cassette gas concentration
agent-specific cassette (Aladin cassette):
-does not contain a bypass channel
-anesthetic liquid cannot escape while handing (ex. filling)
-cassette can be carried in any position without concern of 'tilting' are related concern of resultant hazardous vapor output
-the flow leaving the saturated anesthetic liquid chamber combines with the bypass flow to join before leaving the fresh gas outlet
-altering the ratio between the ratio between the bypass flow and liquid chamber flow changes the concentration of the anesthetic
-sensors in the cassette measure pressure and temperature which helps calculate the exiting anesthetic agent concentration
Ventilators
-create a pressure gradient from the proximal airway to the alveoli
-newer models create positive pressure ventilation (vs old models using negative pressure ex. iron lungs)
Four phases of ventilation:
-inspiration
-transitional period between inspiration to expiration
-expiration
-transitional period between expiration to inspiration
Mechanical Ventilators
generate a tidal volume by producing a pressure gradient which directs as flow
Termination of inspiratory phase occurs by preset limit of:
-time duration ex. Time-cycled ventilators allows variation of tidal volume and peak inspiratory pressures
-inspiratory pressure ex. Pressure-cycled ventilators reached preset pressure allows for the cycle from insp to exp phase
-tidal volume ex. Volume-cycled ventilators preset volume creates a varying inspiration duration and insp pressure
Modes of ventilation:
-controlled ventilation
-assisted ventilation
-intermittent ventilation
-synchronized intermittent mandatory ventilation
Waste Scaveging System
-vented gases from breathing circuits are disposed of by a waste-gas scavenging system
-pressure release valve allows for the transfer of gases from the breathing circuit into the scaveging system
-reducing inhalational anesthetic levels to trace levels accepted by NIOSH is accomplished with a functional scaveging system
-NIOSH = national institute of Occupational Safety and Health
-passive scaveging:scaveging system leading to directly to the outside environment
-active scaveging : scaveging system leading to the hospital's vacuum system
Humidifiers
relative humidity: ratio of mass of water present in a volume of gas
absolute humidity
Upper respiratory tract
-100% relative humidity which is approximately 47 mm H20/L at 37 degree C
-normally warms inhaled gases
Inhaled gases via ETT bypass the upper respiratory tract may cause:
-dehydration of musosa
-altered function of respiratory cillia
-inspissations of respiratory tract secretions
-atelectasis possibly leading to V/Q mismatch
Humidifiers:
-reduce water loss
-reduce heat loss
Designs:
-condenser humidifier
-heat and moisture exchanger
traps exhaled humidification and releases the humidity in the next inhalation
Nebulizers
-suspend water particles into a spray in which the droplet size depends on the methodology of nebulization
Method of nebulizing water droplets:
high-pressure jet nebulizers 5 - 30 um diameter droplets
ultrasonic nebulizers 1- 10 um diameter droplets
Oxygen Analyzer
low-level alarm which is activated when turning on the analyzer
placed in the inspiratory limb or expiratory limb of the circle system breathing circuit
not to placed in the fresh gas flow line
measurement of oxygen concentration can be done as follows:
-electrochemically
-paramagnetic analysis
-mass spectrometry
Electrochemical sensors
-galvanic cell
-polarographic cell
paramagnetic sensors
-self callibrating
-no consumable parts
-rapid response time which allows for differentiation of inspiratory and expiratory oxygen concentrations
Comments
Anesthetic Machine
I want to take this moment to say that I really love this blog.
It has been a good resource of information for me in my research about Anesthetic Machine.
silk blouses | women's blouses | vintage blouses
Thank you so much.
Hi I am so glad I discovered
Hi I am so glad I discovered your site, I truly found anyone by miscalculation,
while My partner and i was seeking on Askjeeve for something else,
Nonetheless My business is here right now and would just as to state cheers for
a incredible post along with a all around enjoyable blog
Resume help
hi
That post is written in very a good manner and it entails much useful information for me. The code looks easy to implement. I appreciated what you have done here. I am always searching for informative information like this. Thanks for sharing with us.Best Wireless Mouse|Logitech Wireless Mouse
Austin TX Home Builder
Thank you for this blog. Thats all I can say. You most definitely have made this blog into something thats eye opening and important. You clearly know so much about the subject, youve covered so many bases.Austin TX Home Builder
Austin TX Home Builder
I suggest this site to my friends so it could be useful & informative for them also. Great effort.Austin TX Home Builder
I like all your post. I will
I like all your post. I will keep visiting this blog very often. It is good to see you website.seattle dentist
steel building
I just wanna thank you for sharing your information and your site or blog this is simple but nice article I have ever seen i like it i learn something today
steel building
steel building
I hope you’ll have the time soon to open this discussion back up. Best regards
steel building
The post is pretty
The post is pretty interesting. I really never thought I could have a good read by this time until I found out this site. I am grateful for the information given.
stargames
Tube -hypoxic mixtures are
Tube -hypoxic mixtures are possible if a leack develops within or downstream of the 02 flowmeter -risk of hypoxic mixture is reduced if the 02 flow meter is positioned downstream
cialis en ligne
cialis en ligne
Great
Fantastic article, this is so well explained that even my kids could follow it, thanks and keep up the great work!
affiliatenätverk
Tube -hypoxic mixtures are
Tube -hypoxic mixtures are possible if a leack develops within or downstream of the 02 flowmeter -risk of hypoxic mixture is reduced if the 02 flow meter is positioned downstream
thesis writing service
I am almost excited with all
I am almost excited with all the information of your site.I really do enjoy this great website and your post was really amazing. Automatische externe defibrillator AED
Wonderful post! I've
Wonderful post! I've bookmarked your weblog as well considering I found it issurely educational and I liked reading your posts.I really enjoyed this wonderful post that you have provided for us. get your ex girlfriend back get back with your ex
outdoor sheds
This is a very intriguing post, I was looking for this knowledge. Just so you know I found your web site when I was searching for blogs like mine.
outdoor sheds
fat loss diet
That is really interesting, You’re a very skilled blogger. I’ve joined your feed and stay up for searching for extra of your great post. Also, I have shared your web site in my social networks. fat loss diet
Fantastic looking site! I
Fantastic looking site! I have a site as well but unfortunately I'm not very good with design and other computer (tech) related stuff, but would like to make my theme a bit more custom as well. Could you refer me to your designer please? Thankviagra generique
Xtreme NO
It is hard to find post of these sort, I am pretty pleased with it and I highly recommend it.
Xtreme NO
Nue Science
acheter viagra
I found some really interesting stuff in your blog especially this discussion. Keep up the good work. acheter viagra
acheter viagra
Your post is rocking and knowledgeable... I really appreciate the way you write . I would like to read more from you.acheter viagra
dentist seattle
Hey, I had been searching on this topic for a long while but I was not able to find great resources like that. Now I feel very confidence by your tips about that, I think you have choosen a great way to write some info on this topic.
dentist seattle
steel buildings
Nice post.Thank you for taking the time to publish this information very useful!Im still waiting for some interesting thoughts from your side in your next post thanks.
steel buildings
Tree Care
Thanks for this post. I definitely agree with what you are saying. I have been talking about this subject a lot lately with my mother so hopefully this will get him to see my point of view. Fingers crossed! tree care
Wood Chipper,Pellet Mill,Air Cooler and Mixers supplier
This is exactly what I was looking for. Thanks for sharing this great article ! That is very interesting Smile I love reading and I am always searching for informative information like this ! hammer mill,Wood Chipper,Air Cooler,Mixers and Pellet Mill supplier
Samuel
The content truly deserves appreciation as it is very unique and seems to be working pretty fine.
Argan oil
30 Second Smile
Additions with regards to
Additions with regards to apparels will be the vests that Moncler offers.goose outletwhile working with exclusive warm and appealing colorations that this vests look in,goose saleits a new handle for that eyesight enchanting that process trend. exactly how you you should get some vest is how you define your design. goose jakkerMoncler guys jackets The actual nylon soluble fiber while from the vest will keep you inviting and Moncler establishes that males are not anxious to check with shades. cottage
I am really comply with this
I am really comply with this article. It is a valuable and charming article for all. Everyone please visit this site and discus about this post each other.
rhinoplasty New York
real estate la jolla
You helped me greatly in my work by writing this article. I want to thank you for your effort. Your site is great.
real estate la jolla
Silk n Sensepil
I am pretty much content with with the entire concept of the post.
Silk n Sensepil
I am almost excited with all
I am almost excited with all the information of your site.I really do enjoy this great website and your post was really amazing. This is the best blog for someone who would like to know about this matter. Thanks for sharing all these important infos. partytent huren
Silk n Sensepil
I am pretty much content with with the entire concept of the post.
Silk n Sensepil
Anesthetic machine is a very
Anesthetic machine is a very helpful for all the medical surgery. I was able to find the details that I was searching for. I must thank you for the initiatives you have made in writing this article. love spells to win your ex back how to win your ex back
Great
This is my first time i visit here. I discovered a lot of interesting things within your blog especially its discussion.
blek tänderna innan semestern
hi
Let me appreciate the thoughts and creativity of the Author of this post.I enjoy prospect behind the complete concept.
बड़ा लिंग
enlarge your penis
beauty salons
The Firefox SEO extension looks amazing, but when I updated my version of Firefox, the extension that’s available for download no longer works. I use the SEO extension for Googlehair salons
hi
I most certainly will appreciate the thoughts and ingenuity of the Author of this post.I enjoy prospect behind the full concept.
बड़ा लिंग
enlarge your penis
hair salons
My partner and i fluctuate with a lot of folks right here; I came across this website article My spouse and i couldn’t leave until finally My partner and i concluded, eventhough it wasn’t just what I used to be trying to find, had been a pleasant go through however.beauty salons
Atlanta Property Management
Happy to see your blog as it is just what I've looking for and excited to read all the posts. I am looking forward to another great article from you. Siam@atlanta property management
You’re exactly correct, I
You’re exactly correct, I think using that phrase, even among native speakers, you may not be immediately understood. It’s something that is relatively uncommon and mostly used in writing. portugese tegels
Baby Bullet
Without using it many things will not happen as a result we have to use it.
Baby Bullet
Thanks
Thank you for article you've written. It will eventually give me what to dream about while I go to florida vacation
I most certainly will
I most certainly will appreciate the minds and creativity of the Writer of this post.I love thinking behind the complete subject.
बड़ा लिंग
enlarge your penis
Makeityourring Diamond Engagement Rings
You got numerous positive points there. I made a search on the issue and found nearly all peoples will agree with your blog.
Makeityourring Diamond Engagement Rings
My cousin is asthmatic and I
My cousin is asthmatic and I often see him use a nebulizer. I didn't know that there were a variety of vaporizers available. I wonder how they are classified to patients.
credit score ratings
Great!!!
we have to protect this blogpost and the comments against spam. anyway its good visit for me...
phlebotomist
Thanks
Welcome to cheap COACH store.coach outlet When you think about all of the positives offered by this Everything about a coach handbag outlet is unique, right down to the The first and larger location has over over 200 stores for Orlando shopping malls are a theme Nike Free Runs park experience for adults.Berufstest Berufswahltest
Anesthetic machine is a very
Anesthetic machine is a very helpful for all the medical surgery. I was able to find the details that I was searching for. I must thank you for the initiatives you have made in writing this article. Augmented Reality
Thanks
The anaesthetic machine is used by anaesthesiologists and nurse anaesthetists to support the administration of anaesthesia. The most common type of anaesthetic machine in use in the developed world is the continuous-flow anaesthetic machine.cold sores on lips |adirondack chair plans
I’m in Florida but I will
I’m in Florida but I will be in MA April 20th – 25th and I’d be really interested in doing a short ride.I’m in Florida but I will really interested in doing a short ride.Bandera Homes
Helotes Real Estate
Thanks for very interesting
Thanks for very interesting post. I have a high regard for the valuable information
you offer in your articles. I really believe you will do much better in the future.
cialis